Skip to content
English
  • There are no suggestions because the search field is empty.

Vista Touch Isolation Reference

Isolation prevents electrical noise, surges, or ground faults in one part of a system from affecting other parts. This improves safety, measurement accuracy, and overall reliability. The table below shows which inputs and outputs are electrically isolated for Trumeter Vista Touch meters, so you can quickly see how each product will behave in your installation.

Vista Touch Meter Power Input Voltage Input Current Input Digital Inputs Digital/Relay Outputs
Vista Touch Flow Isolated N/A N/A Isolated Isolated (Relay)
Vista Touch Power N/A Isolated (not isolated from current input) Isolated (not isolated from voltage input) Isolated Isolated (Relay)